Ceramic Multigen Plus Knee With Biolox Delta Ceramic Femoral Component

Ceramic Multigen Plus Knee With Biolox Delta Ceramic Femoral Component Download Scientific Diagram

Ceramic Multigen Plus Knee With Biolox Delta Ceramic Femoral Component Download Scientific Diagram

Ceramic Femoral Components In Total Knee Arthroplasty Two Year Follow Up Results Of An International Prospective Multi Centre Study

Ceramic Femoral Components In Total Knee Arthroplasty Two Year Follow Up Results Of An International Prospective Multi Centre Study

Fixed Or Mobile Bearing Knee Prosthesis Multigen Plus Cr Multigen Plus Ps Lima Corporate Cemented Or Non Cemented

Fixed Or Mobile Bearing Knee Prosthesis Multigen Plus Cr Multigen Plus Ps Lima Corporate Cemented Or Non Cemented

Ceramic Femoral Components In Total Knee Arthroplasty Two Year Follow Up Results Of An International Prospective Multi Centre Study

Ceramic Femoral Components In Total Knee Arthroplasty Two Year Follow Up Results Of An International Prospective Multi Centre Study

Http Www Ceramtec Com Cn Files Mt Ceramic Knee En Pdf

Http Www Ceramtec Com Cn Files Mt Ceramic Knee En Pdf

Https Journals Sagepub Com Doi Pdf 10 1177 230949901302100306

Https Journals Sagepub Com Doi Pdf 10 1177 230949901302100306

Https Journals Sagepub Com Doi Pdf 10 1177 230949901302100306

Ceramic multigen plus knee with biolox.

Ceramic multigen plus knee with biolox delta ceramic femoral component.

The use of ceramic femoral components in total knee replacement provides tribological and allergological benefits compared to metallic components. The multigen plus total knee system lima corporate villanova di san daniele del friuli italy is a non constrained surface replacement consisting of a symmetric cruciate retaining cr cemented biolox delta ceramic femoral component the cemented metal tial6v4 tibial tray is combined with fixed bearing ultra high molecular weight polyethylene liners uhmwpe. Its design is identical to that of the multigen plus knee except for having a symmetric anterior surface. All tkrs were implanted in a standardised.

The clinical outcome of the multigen plus knee with a biolox delta ceramic femoral component demonstrates results comparable to other studies which reported a mean postoperative 1 year hss score between 85 0 to 93 0 points 20 21. The first implantation of the multigen plus knees with the biolox delta ceramic femoral component was successfully performed in our department in december 2006. The multigen plus system with a ceramic femoral component lima corporate villanova di san daniele udine italy consists of a biolox delta ceramic femoral component. Clinical and radiological evaluations were performed preoperatively and after 3 12 and 24 months.

A national prospective duocenter study on the clinical and radiological outcome of the multigen plus total knee system with a biolox delta ceramic femoral component article january 2010 with 17. The multigen plus ceramic knee is a non constrained. The ceramic components receive precisely executed final polishing which results in reduced surface roughness around 2 nanometers of all implant materials. Facing the brittleness and the low.

Major prerequisite for the development of the biolox s delta ceramic knee components was the use of existing designs in order to provide the surgeon with a well known surgical technique and the same instruments as for the standard metal femoral component short learning curve. Biolox delta ceramic femoral component integrated in the multigen plus total knee system lima lto san daniele italy is introduced in the european market. The study was aimed to evaluate the clinical and radiological outcomes of the biolox delta ceramic femoral component and was classified as evidence. Cemented total knee arthroplasty was performed on 108 patients 110 knees at seven hospitals in three countries.

A total of 25 tkas were included in a prospective clinical study in agreement with the inclusion and exclusion criteria after approval of the local ethical committee.

Bioloxdelta Components For Tka Lower Left Femoral Component Of The Download Scientific Diagram

Bioloxdelta Components For Tka Lower Left Femoral Component Of The Download Scientific Diagram

Bpk S Integration Total Knee Replacement System With Femoral And Tibial Download Scientific Diagram

Bpk S Integration Total Knee Replacement System With Femoral And Tibial Download Scientific Diagram

The All Ceramic Knee Endoprosthesis The Gap Between Expectation And Experience With Ceramic Implants Sciencedirect

The All Ceramic Knee Endoprosthesis The Gap Between Expectation And Experience With Ceramic Implants Sciencedirect

Comparison Of Selected Properties Between Contemporary Medical Grade Download Table

Comparison Of Selected Properties Between Contemporary Medical Grade Download Table

Source : pinterest.com